Skip to content

Phesgo Compatibility and Administration by using infusion sets or syringe pumps

This article responds to your request for information on Phesgo® (pertuzumab, trastuzumab, and hyaluronidase) subcutaneous injection and administration using infusion sets or syringe pumps.

 

Please refer to the locally approved Phesgo prescribing information for recommendations on administering Phesgo. Any deviation from this information is considered off-label, and any treatment decisions based on such deviations are the full responsibility of the prescribing physician. 

Download article

Last updated July 27, 2026

Summary

  • Phesgo prescribing information does not provide details on administering Phesgo with a subcutaneous infusion set, such as a butterfly or winged needle. It only lists materials compatible with Phesgo.
    • Phesgo was not developed to enable the use of subcutaneous infusion sets or syringe pumps. Roche cannot provide guidance on the practical use of infusion sets or pumps, and the decision to use a subcutaneous infusion set is at the discretion of the treating healthcare professionals (HCPs).
      • Be aware that infusion sets can retain some of the drug product, known as hold-up volume.
        • Use of an infusion set
          • The only infusion set tested by Roche that fully meets the product quality and acceptable hold-up volume requirements for Phesgo is the Ypsomed mylife YpsoPump Orbit soft 6mm/60cm or 45cm.
          • Use of syringe pumps
            • In a technical feasibility study, Roche tested the administration of Phesgo with syringe pumps and winged infusion sets using butterfly needles. Roche found that the B. Braun Perfusor Space and Terumo Terufusion TE-SS800 syringe pumps were suitable for automated flow rate control of Phesgo without triggering the occlusion alarm. The Ypsomed mylife Ypsomed Pump Orbit soft 6mm/60cm infusion set was used with each pump.

            Compatibility of materials and Phesgo

            No incompatibilities between Phesgo and polypropylene (PP), polycarbonate (PC), polyurethane (PU), polyethylene (PE), polyvinyl chloride (PVC) and fluorinated ethylene polypropylene (FEP) have been observed.[1] 

            Administration using a butterfly or winged needle

            Roche has no recommendations on administering Phesgo using a butterfly or winged needle. Refer to the locally approved prescribing information for further recommendations on appropriately sized needles for the administration of Phesgo.

            Use of infusion sets

            Roche has no recommendations on administering Phesgo with a subcutaneous infusion set.[1] The decision to use a subcutaneous infusion set is at the discretion of the treating healthcare professionals (HCPs). Refer to the locally approved prescribing information for further recommendations on appropriately sized needles for the administration of Phesgo.

            In-use compatibility

            The only infusion set tested by Roche that meets material compatibility with an acceptable hold-up volume requirements for Phesgo is the Ypsomed mylife YpsoPump Orbit soft 6mm/60cm or the 45cm version.[2] Using infusion sets with materials not mentioned in the Phesgo label may impact product quality and pose a potential safety risk to patients.

            Hold-up volume

            Infusion sets or systems can retain some of the drug product, known as hold-up volume. Phesgo vial fill volumes were not developed to enable the use of subcutaneous infusion sets. Healthcare professionals are responsible for ensuring the entire dose of Phesgo is administered to the patient.

            An internal technical study assessed the hold-up volume of the Ypsomed mylife YpsoPump Orbit soft 6mm/60cm infusion set when used to administer Phesgo.[3] The study found that the Ypsomed mylife YpsoPump Orbit soft 6mm/60cm allowed delivery of the entire dose of Phesgo, after withdrawal from the vial using a syringe and transfer needle. The same set, but 45cm tubing would also be considered acceptable for Phesgo.

            Use of syringe drivers

            Roche has no recommendations regarding the administration of Phesgo with a syringe driver or pump. The decision to use a syringe driver is at the discretion of the treating healthcare professionals (HCPs).[1]

            Occlusion alarm study for two electronic syringe pumps

            In an experimental pressure characterisation study, Roche simulated the administration of a Phesgo 10 mL solution with two syringe drivers [4]:

            • B. Braun Perfusor Space — Reference number 8713030
              • Terumo Terufusion — Reference number TE-SS800EN1
                 

                Both syringe pumps were tested with Ypsomed Orbit SC infusion sets. Back pressure of the skin was not considered in the measurements, with the test simulating injection into the air. Back pressure is thought to be minimal with Phesgo due to the inclusion of recombinant human hyaluronidase. The simulation found that both syringe pumps were suitable for automated flow rate control of Phesgo while using tested infusion sets, without triggering the occlusion alarm.[4]

                References

                1. Roche Internal Regulatory Report (Phesgo CDS 5.0).
                  1. Roche Internal Technical Report (TEC-0172225).
                    1. Roche Internal Technical Report (TEC-0173596).
                      1. Roche Internal Technical Report (2018-363C).

                        Welcome to Medically

                        The Roche Science Hub

                        This website is a non-promotional global resource intended to facilitate transparent scientific exchange regarding developments in medical research, diagnostics, and disease management.

                        Not a healthcare professional? Browse:

                        This website is a non-promotional global resource intended to facilitate transparent scientific exchange regarding developments in medical research, diagnostics, and disease management. This global website is intended for healthcare professionals outside the UK, US, Canada, and Australia. The content on this website may include scientific information about experimental or investigational compounds, indications, and services that are not approved or valid in your country. Registration status and prescribing information of medicinal products may differ between countries. Please refer to local product information for any medicinal products mentioned on this website. Information available on this website does not constitute professional medical advice, and Roche and Genentech accept no responsibility for access to or use of the same.

                        You are Leaving Medically

                        By following this link, you are leaving Roche Website and entering a site that is not owned or controlled by Roche. Roche does not take any responsibility for acces to or use of this website, nor for any content therein.

                        Leave Site

                        You are Leaving the Global Medically Site

                        By following this link, you are being redirected to another Roche page.